Features and Benefits
•Measures CO2, temperature, humidity, and CO
•Calculates dew point, wet bulb and percent outside air
•Displays up to 5 measurements simultaneously
•One instrument with multiple plug-in probe options
•User-selectable logging intervals and start/stop times
•Store up to 38.9 days of data collected at one-minute log intervals
•TRAKPRO™ software provided for data logging, analysis and documenting results
•Perform single-point tests for quick building surveys
Applications
•Measure indoor air quality conditions in schools, offices and other indoor environments
•Maximize thermal comfort for optimum productivity
•Investigate Sick Building Syndrome and building related illness
Specifications
Specifications are subject to change without notice.
CO2:
Range: 0 to 5000 ppm
Accuracy1: ±3% of reading or ±50 ppm, whichever is greater
Resolution: 1 ppm
Sensor type: Non-Dispersive Infrared (NDIR)
Temperature:
Range: 32 to 140°F (0 to 60°C)
Accuracy: ±1.0°F (±0.6°C)
Resolution: 0.1°F (0.1°C)
Response time: 30 seconds (90% of final value, air velocity at 400 ft/min
[2 m/s])
Type: Thermistor
Relative Humidity:
Range: 5 to 95% RH
Accuracy2: ±3% RH (includes ±1% hysteresis.)
Resolution: 0.1% RH
Response time: 20 seconds (for 63% of final value)
Sensor type: Thin-film capacitive
CO Sensor:
Range: 0 to 500 ppm
Accuracy: ±3% of reading or 3 ppm whichever is greater [add
±0.5%/°C (0.28%/°F) away from calibration temperature]
Resolution: 1 ppm
Response time: <60 seconds to 90% of final value.
Sensor type: Electro-chemical
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Storage: -4 to 146°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
Data Storage Capabilities:
Range: Logs up to 56,035 data points with key (4) measured
parameters enabled, 38.9 days at 1-minute log intervals
Logging Interval:
Intervals: 1 second to 1 hour (user selectable)
Time Constant:
Intervals: 1 sec, 5 sec, 10 sec, 20 sec, 30 sec, (user selectable)
External Meter Dimensions:
3.8 in. x8.3 in. x2.1 in. (9.7 cm x 21.1 cm x5.3 cm)
Meter Probe Dimensions:
Probe length: 7.0 in. (17.8 cm)
Probe diameter of tip: 0.75 in. (1.9 cm)
Meter Weight:
Weight with batteries: 0.8 lbs (0.36 kg)
Power Requirements:
Batteries: Four AA-size alkaline or rechargeable
or
AC adapter: 6 VDC nominal, 300 mA [Q-TRAK monitor mates with
5.5 mm OD x2.1 mm ID plug, center pin positive(+)]
Easy to Use
The Q-TRAK features a menu-driven user interface for easy operation. On-screen prompts and step-by-step instructions guide the user through operation and field calibration. The Q-TRAK also features an ergonomic, overmolded case design and a keypad lockout to prevent tampering during unattended use.
Accurate Results
The Q-TRAK monitor’s state-of-the-art sensors and large, easyto-read graphics allow real-time, simultaneous display of CO2 temperature, relative humidity, and CO.
Data Collection and Reporting
Expanded data logging capacity and the inclusion of TRAKPRO Data Analysis Software provides the capabilities to work more effectively and efficiently. The Q-TRAK can store up to 38.9 days of data collected at one-minute log intervals. The stored data can be recalled, reviewed on screen, and downloaded for easy reporting. TRAKPRO helps you to generate professional graphs for your reports.
Use the request button on product pages to build a quote request list.
| 0 Items | $0.00 |