Envco – Environmental Equipment

User login

Hydro Services WL2100D Dry Pressure Transducer

The Hydrological Services WL2100D is our latest generation of dry pressure transducers. It allows the measurement of water head to a fine degree of accuracy and repeatability, due to full temperature compensation, and interfaces via SDI-12 or 4-20mA.
Hydrological Services WL2100D Dry Pressure Transducer
Hydrological Services WL2100D Dry Pressure Transducer

The Dry transducer consists of a strain gauge bridge sensing element fitted to a housing. The strain gauge bridge is type 316 Stainless Steel and the housing is a power coated die-cast enclosure. The electrical connection is made via a multicore cable .

The WL2100D output is both SDI-12 and 4-20mA compatible but not at the same time. This allows any SDI-12 or any 4-20mA recording device to communicate with the pressure transducer. The operator may use the SDI-12 to RS232 adaptor cable and then communicate using RS232 from a PC. This special mode is enacted on each command by proceeding each SDI-12 command with an ascii ‘*’ rather than a “break’. When the ascii ‘*’ is detected, all of the timing/break requirements of the SDI-12 are removed. Commands may be typed from a terminal program such as “Hyperterm”.

Operating Range : 100kPa (10 metres), 200kPa 20m), 300kPa (30m)
Overpressure : 1.5 x Range
Supply Voltage : 9V – 30VDC (beyond SDI-12 specification Version 1.2)
Supply Current : 2.0mA nominal in SDI-12 mode

Output Signal : SDI-12 Compliant (Version 1.2) OR 4-20mA (The user selects)
Overall Accuracy : + 0.05% F.S. BSL
Long term stability : Typically + 0.03% FS/annum

Measurement Cycle : Every 4 secs (in 4-20mA mode), When requested (in SDI-12 mode)
Recommended Warm Up Time : 15 secs (after power is first applied)
Operating Temperature range : -20C to +60C (Compensated over this range)

Electrical Connection : 3 Wires +12V Power (red) SDI-12 Data (white) 0V Gnd (black)
Vented Cable : 1 metre of 3 core vented cable
Voltage Spike Protection : Will withstand a 600V voltage spike in accordance with ENV50142 with out damage when applied between excitation lines and case.
Calibration : Over full pressure range.
Insulation : Greater than 100Mohm at 500V d.c.

Dimensions : Length 150mm
Width 60mm
Height 35mm
Mass : 0.45 kg (excluding cable)

Connection as an SDI-12 Transducer
**Note that the WL2100 is supplied set to the 4-20mA mode as default. Therefore to operate in SDI-12 mode with minimal power consumption, the X14 parameter must be set to 0. That is 0X14+0!

Ordering Information
WL2100W - SUBMERSIBLE PRESSURE SENSOR +/- 0.05% FS ACCURACY, ‘FULLY’
TEMPERATURE COMPENSATED, ‘DELRIN’ HOUSING, NICKEL PLATED BRASS
NOSE CONE (DELRIN NOSE CONE AVAILABLE ON REQUEST), OUTPUT SIGNALS
4-20mA OR SDI-12. INCLUDES JUNCTION BOX, COMPLETE WITH SCINTERED
FILTER
RANGES - WATER LEVEL
................................. 0-5M,
................................. 0-10M,
..................................0-15M,
..................................0-20M,
..................................0-30M
OTHER RANGES AVAILABLE UPON REQUEST

If you cannot find the part number in the documentation please provide detail. Specific to this item, it could include part number or application detail, (ex. 20 meters of cable, or 20 litres of volume)